Operational sequence
Measured value display (measurement mode)
Entry into the main menu
Entry into a
sub-menu
FIDAMAT 6 gas analyzer
Operating manual –A5E00222135-01
Attention
To prevent static charges the keyboard should only be used for
maintenance or operating purposes.
The measurement component, furnished with a rightward pointing arrow
(►), appears on the right-hand side of the screen. A soft key is assigned
to this component. Pressing this soft key calls the main menu.
The main menu consists of the following sub-sections (sub-menus).
To their
right is shown the corresponding code level:
Diagnosis
Calibration
Measuring ranges
Parameters
Configuration
The code for level 1 is factory-set to "111" and that for level 2 is set to
"222."
Once a sub-menu has been selected, you will be asked for the code for
the operator level (exception: the "Diagnosis" submenu is accessible
without restrictions).
Decoding level 2 also decodes level 1.
With decoding, signaling to the outside can take place via a relay contact
provided a corresponding relay has been configured via function 71 with
CTRL. The warming-up and calibration phases of the unit are also
signaled via this relay contact.
With decoding the measured value memory is also activated, provided it
was set under function 77. The coding of the unit is to be seen on the
Measured value display) as the symbol
display (Fig.:
decoding with
CODE.
Not coded
Code level 1
Code level 1
Code level 1
Code level 2
Operation
CODE,
5-5