Siemens FIDAMAT 6 Operating Instructions Manual page 17

Gas analyzer for the measurement of total hydrocarbons
Table of Contents

Advertisement

Main features
Four measuring ranges freely
parametrizable, also with suppressed
zero point, all ranges linear
Hydrogen required: H 2
Extremely short measuring spans
Electrically isolated measured-value
output 0/ 2/ 4 to 20 mA (negated also)
Automatic or manual range changeover
selectable; remote changeover is also
possible
Measured values can be retained during
calibration
Time constants selectable within wide
limits (static / dynamic noise
suppression); in other words, the Unit's
response time can be adjusted to suit the
measurement task on hand
Simple to use due to menu-controlled
operation
User operation based on NAMUR
recommendations
Short response time
Low long-term drift
Three operator control levels each with
own authorization code to prevent
inadvertent or unauthorized operator
actions
FIDAMAT 6 gas analyzer
Operating manual –A5E00222135-01
External pressure pick-ups can be
connected to correct process gas
pressure fluctuations
Units easy to swap over since electrical
connections only need to be
disconnected from unit
Wear-free, corrosion-resistant filter
No clogging of measuring-gas capillary
since a quartz capillary is used
Low consumption of combustion air
Digital display of gas concentration, of
measuring gas, combustion air and
hydrogen pressure
Response factors meet the minimum
requirements of the Clean Air Guidelines
and of the German Automotive Industry
Working Group
Simplicity of operation due to a numerical
membrane keyboard and operator
guidance
Parametrizable automatic measuring
range calibration
PROFIBUS (DP/ -PA) / AK
Menu and interface are compatible with
the other Series 6 devices (CALOMAT 6,
OXYMAT 6, ULTRAMAT 6, OXYMAT 61)
Customer-specific versions of the device,
such as:
- Customer acceptance
- TAG plates
- Drift recording
Technical description
2-3