Technical specifications
9.2 General technical specifications
Table 9-18
OXYMAT 7: Climatic conditions
Climatic conditions
Storage and transport
Max. permissible ambient temperature surrounding the basic
device during operation
1)
Ambient humidity (rel. humidity) during storage, transport or
operation
2)
Restriction for installing an ULTRAMAT 7 analyzer module (combined operation): 5 ... 45 °C (41 ... 113 °F)
1)
Avoid condensation on the installed components!
2)
Table 9-19
OXYMAT 7: Gas connections
Gas connections
With pipes
Table 9-20
OXYMAT 7: Parts in contact with sample gases
Materials of parts in contact with
sample gases
Sample chamber
Gas path
Sealing material
1)
The high-temperature version of the analyzer module is not intended for installation in the rack unit.
9.2.5

Technical data ULTRAMAT 7

Table 9-21
ULTRAMAT 7: General technical specifications
General
Information on the smallest span
Max. power consumption
90
- 30 ... 70 °C (-22 ... 158 °F)
0 ... 50 °C (0 ... 122 °F)
< 90%
Connecting socket for pipe with ∅ 6 mm
OXYMAT 7
Stainless steel:
Plates: Material No. 1.4571
Screw-in glands: No. 1.4571
Hastelloy:
Plates: Material No. 2.4602
Screw-in glands: Material No. 2.4819
Stainless steel:
Pipes: Material No. 1.4404
Gas connections: Material No. 1.4571
Hastelloy:
Pipes: Material No. 2.4602
Gas connections: Material No. 2.4819
FKM (e. g. Viton) or FFKM (Kalrez 6375)
ULTRAMAT 7
≤ 30 W
OXYMAT 7 high temperature
Stainless steel:
Plates: Material No. 1.4571
Hastelloy:
Plates: Material No. 2.4602
Stainless steel:
Gas connections: Material No. 1.4571
Hastelloy:
Gas connections: Material No. 2.4602
ULTRAMAT 7 high-temperature
See module nameplate
≤ 120 W
Devices in explosion-proof design
Compact Operating Instructions, 01/2022, A5E35134047-11