Siemens FIDAMAT 6 Operating Instructions Manual page 85

Gas analyzer for the measurement of total hydrocarbons
Table of Contents

Advertisement

The next diagram gives an overview of the interactions between the various memory modules.
Fig. 5-5: Memory modules
76 Suppression of
short-duration fault signals
FIDAMAT 6 gas analyzer
Operating manual –A5E00222135-01
This function is used for getting rid of unwanted steep-edged faults which
exceed a configurable threshold.
Steep-edged faults ("spikes") are caused by electromagnetic interference
or occasional mechanical jolts. These faults can be suppressed by
inputting a "blanking-out time" of 0...5 s. The time input causes the last
measured value before a spike occurred to be output and the
measurement result is thus no longer affected.
Input can be made in steps of 0.1 s.
If a change in concentration follows immediately upon a fault, this change
may under certain circumstances be displayed after a delay.
When activating this function, the settings for function 50 ("Electrical time
constants") must be taken into account since this function is executed
first.
Operation
5-35