Example:
The following sequence is to be programmed:
1. Zero gas calibration, with zero gas flushing after 15 minutes
2. Calibration with test gas 1, with flushing with test gas after
10 minutes
3. Flushing with sample: 8 minutes
4. Sample gas intermediate operation: 30 minutes
5. Calibration with test gas 2, with flushing after 8 minutes
6. Calibration with test gas 3, with flushing after 8 minutes
7. Calibration with test gas 4, with flushing after 10 minutes
8. Flushing with sample: 8 minutes
9. Momentary signaling contact so as to be able to start
"AUTOCAL" in another unit or channel.
Note
The FIDAMAT 6 only has one connection for test gas. Additional test
gases will need to be connected up externally and switched using
external valves.
The AUTOCAL sequence specified is shown in the screen to the left.
List for the AUTOCAL sequence:
Step
Zero gas 1
Zero gas 2
Test gas 1
Test gas 2
Test gas 3
Test gas 4
Flush sample
Sample intermediate
operation
Signaling contact
Zero gas 1b
Test gas 1b
Test gas 2b
Test gas 3b
Test gas 4b
Component
Component 1
– " –
– " –
– " –
– " –
– " –
Component 2
– " –
– " –
– " –
– " –
Operation