57 Ignite flame
58 Date and time
FIDAMAT 6 gas analyzer
Operating manual –A5E00222135-01
To get the FID running the hydrogen flame must be ignited.
This is done via function 57.
Ignition is activated
or deactivated
In addition, the temperature of the flame and of the sample stream are
shown on the display.
When the flame is burning, "Flame burning" appears on the display. If the
flame goes out, this message disappears.
With this function you can set the time and date precisely.
When the function is called, the presets shown in the screen shot on the
left appear in the display.
The cursor will already be standing on the first position of the date display
which can be changed in the order day, month, year. The same procedure
applies to the time (hours, minutes).
The unit has a system clock which is not protected against power failure
(not a real-time clock).
Attention
When switched off, the clock remains at a standstill and is not updated.
The main importance of the settings is for trouble-shooting. With the aid of
the time and date it is easier to classify faults which have occurred and
which are always recorded in the logbook (function 60).
By pressing the third soft key ("Set clock") the data which have been set
are taken over for use. They will then appear at the bottom edge of the
screen.
by pressing the second hot key.