Diagnosis - Siemens FIDAMAT 6 Operating Instructions Manual

Gas analyzer for the measurement of total hydrocarbons
Table of Contents

Advertisement

Operation

5.3.1 Diagnosis

1 Factory data
2 Diagnostic values
3 Logbook
5-10
When the diagnostic functions are selected in the main menu by pressing
the first soft key ("Diagnosis") the adjacemt display appears.
Access to diagnostic functions is not restricted. For this reason you will
not be asked for the code.
Every unit provides the following diagnostic functions:
When this function is selected, important manufacturer's data for the unit
are displayed:
Motherboard:
Firmware no.:
Order number of the software held in the EPROM for the
motherboard
Order no.:
Information about ordering data for the unit
Serial no.:
Information about date of manufacture and unit serial number
Article status:
Information about hardware configuration of the unit
Software status and date
Information about the unit's functional scope at the motherboard
The most important diagnostic values are listed under function 2. They
may allow conclusions to be drawn for evaluating setting activities or
errors.
In the logbook will be found a list of all errors which resulted in a
maintenance request (W) or a fault message (S). (see also on page 6-5
"List of maintenance requests")
Limit alarm (LIM) and function check (CTRL) are also recorded.
These, however, do not trigger a maintenance request or fault message.
The logbook contains a maximum of eight pages, each of which can hold
four messages. It works on the principle of the cyclic buffer – in other
words, when all eight pages (all 32 slots) are occupied, the oldest
message will be overwritten.
Logbook entries can be deleted or locked (function 60) and even switched
off individually (function 87).
Note
If an error occurs and its error message has been switched off with
function 87, there will be no reaction at an interface which may have been
configured. This applies to the ELAN interface and also to the analog
output and relay output.