Prior Scientific ProScan III Manual page 74

Universal microscope automation controller
Hide thumbs Also See for ProScan III:
Table of Contents

Advertisement

ProScan® III Manual
Example 1:
Create a Trigger that will be activated when TTL1 goes low and moves the stage and then
open shutter1 on the next rising edge of that input.
TTLTP,1,0
TTLACT,1,40,5000,4000,0
TTLTP,1,1
TTLACT,1,60,0,0,0
TTLTRG,1
First falling edge detected will execute the stage command, the next rising edge of TTL1
causes the shutter to open.
Example 2:
The opening of the shutter can be automated by wiring the output of the TTLMOT
to another TTL input.
TTLTP,1,0
TTLACT,1,40,5000,4000,0
TTL 1. TTLTP,2,1
TTLACT,2,60,0,0,0
TTLMOT,2,1
TTLTRG,1
© Prior Scientific Instruments Ltd
Adds trigger point on TTL 1 on falling edge.
Adds stage movement to current trigger point on TTL 1.
Adds a new trigger point TTL 1 on rising edge.
Add open shutter to current trigger point on TTL 1.
Sets system ready.
Adds trigger point on TTL 1 on falling edge.
Adds stage movement to current trigger point on
Adds new trigger point on TTL 2 rising edge
Add open shutter to current trigger point on TTL 2.
Set the TTL output on TTL 2 to high when stage moving
Set system ready.