Prior Scientific ProScan III Manual page 64

Universal microscope automation controller
Hide thumbs Also See for ProScan III:
Table of Contents

Advertisement

ProScan® III Manual
The command has the following format:
Command
BUTTON
There are 4 possible conditions identified by b
b=0 Button released.(i.e. changing to a state where no button is pressed)
b=1 Right button pressed
b=2 Left button pressed
b=3 'Fire' button pressed (on top of joystick).
(Pressing more than one button simultaneously will not register a command.)
F (a number) defines what action the controller does when it first detects a new state
given by b above. The table below shows which values of f correspond to which actions.
0 Default Button Function
1 Z motor velocity stop (only used with f=2/3 below)
2 Z motor constant velocity up
3 Z motor constant velocity down
4 Toggles digipot (Z-motor) speed (100%/50%/25%)
5 Toggles XY (stage) joystick speed (100%/50%/25%)
6 Toggles joystick speed (100%/10%)
7 Toggles joystick speed (100%/2%)
8 Moves Z motor up by amount defined by C command
9 Moves Z motor down by amount defined by C command
10 Moves stage left defined by X command
11 Moves stage right defined by X command
12 Moves stage back defined by X command
13 Moves stage front by amount defined by X command
14 Toggles digipot speed (100%/10%)
15 Toggles digipot speed (100%/2%)
16 Starts auto focus
23 Next point
24 Next wafer
25 Smooth stop (I)
26 Sudden stop (k)
28 Toggle shutter 1
29 Toggle shutter 2
30 Toggle shutter 3
35 Assigns any following txt to the button. Hence any command may be assignable to a button).
© Prior Scientific Instruments Ltd
Arguments
Response (including <cr>)
b,f
0
Version 1.12 June 2017