ProScan® III Manual

4.8 Pattern Commands

These commands control the internal patterns available in the ProScan controller
Firmware.
Command
Arguments
E
None
E
b
N
n,m
N
None
S
None
S
?
S
n,m
X
x,y
X
None
Y
None
Y
?
Y
n,m
Q
None
T
None
T
?
W
d
W
None
CC
d, r, s
© Prior Scientific Instruments Ltd
Response
Description
Sets the origin for the pattern to the current position. Zeros the
R
pattern X and Y counter.
If b = 0 the stage returns to the origin after completing the final move
0
of the pattern (default setting upon power up). If b = 1 the stage will
stay at the final pattern position.
0
Defines the number n of X and number m of Y steps for the pattern.
n,m
Reports number of X and Y steps.
R
Move to next position in Rectangular Raster.
s
Step number of scan.
R
Moves to nth cell in X and mth cell in Y in rectangular raster.
0
Sets the step size in X and Y
x,y
Reports step size in X and Y.
R
Move to next position of Rectangular Snake.
s
Step number of scan.
R
Moves to nth cell in X and mth cell in Y in rectangular snake.
Sets the origin of the disc grid equal to the current position. Zeros
0
the disc grid X and Y counter.
R
Move to next field of Disc Raster Scan.
s
Step number s of circular scan
0 in
compatibility
mode.
Sets the diameter in mm of a circular disc (Range 1 to 327mm).
Number of
The parameter X defining the step sizes in x and y should be defined
fields in
before setting the W value.
standard
mode.
d
Reports the disc diameter (in mm.)
Where d=diameter (in user units), 100 to 100000
r=number of revolutions, 1 to 1000
s=number of sides to polygon, 1 to 360
The stage then moves backwards to the top of the polygon and
proceeds to move around the side of the polygon sides until all sides
are done.
Version 1.12 June 2017
Page 56