4.18 TTL Programming Advanced Features

More complex features are available that provide the user with the ability to construct
lists of actions to be executed when TTL input lines change state. There is one list per
TTL input. Each list contains one or more trigger points and each trigger point has
associated with it one or more actions. A trigger can be activated by a rising or falling
edge of the TTL input.
Command
n,m
where n = TTL input 1,2,3 or 4 where
TTLTP
m = 0 falling edge or low
m = 1 rising edge or high
TTLACT
n,action,data1,data2,data3
n
TTLDEL
where n = TTL input 1,2,3 or 4
n
TTLRES
where n = TTL input 1,2,3 or 4
M
TTLTRG
where m = 0 Disarmed, 1 Armed
n,m
TTLMOT
where n = TTL output 1,2,3 or 4
where m = 0 low, 1 high
© Prior Scientific Instruments Ltd
Arguments
Version 1.12 June 2017
Response
Adds a trigger point to the TTL input n,
0
activated by a change in the level m
Add an action to the current trigger point
0
for the TTL input n. See table following for
full list of actions.
Deletes all the trigger points and actions
0
associated with TTL input n
Resets the Trigger list for input n back to
0
the start.
0
Arms the global trigger flag
Changes output n to level m while stage is
0
moving. When n=0 function disabled
(default)
Description