ProScan® III Manual
Command
Arguments
SWLL
a
SWLH
a
SWLC
a
VS
x,y
© Prior Scientific Instruments Ltd
Response
Description
Sets the current position of the axis as the low software limit
E.g. SWLL, Z sets current Z position as lowest numerical point for Z
Axes are identified as:
1 or 'X' = X axis
2 or 'Y' = Y axis
0
3 or 'Z' = Z axis
4 or 'A' = fourth axis (linear type only)
NOTES: setting of software limits must be done following a power on and
are set with respect to current XD/YD/ZD settings
Sets the current position of the axis as the high software limit
E.g.
SWLH, Z sets highest numerical point for Z
0
Axes are identified as above
Clears all soft limits on axis a
0
Move the stage at constant velocity in X,Y as specified by parameters. Units
are in microns/s.
R
VS,0,0 will bring the stage to rest. This can be used to simulate a joystick
operation from your application.
Version 1.12 June 2017
Page 50