Lumen Pro Commands

Command
Arguments
LIGHT
LIGHT
n
LIGHT
P,n
LIGHT
P, ?
LIGHT
POWER , n
© Prior Scientific Instruments Ltd
Response
Description
Reports the output of light from the shutter in %.
a
The command automatically locates LGG_SHUTTER. Error 20
reported if no shutter detected.
Sets the output from the shutter to n, where n can be set
between 1-100%.
0
The command automatically locates LGG_SHUTTER. If n is "h"
the shutter will perform a home routine.
Sets the filter wheel position P to n % light output. i.e.,
LIGHT,4,45
Sets position 4 to 45% light output.
Use 7,n,4 to move filter to position 4, where n is the Filter
0
Wheel port which the shutter is attached. STANDARD 10
Position Shutter settings are:
Position %
0,11, 22, 33, 44, 55, 66, 77, 88, 100
n
Reports n, the %output of position p.
Switches on or off the Lumen 200PRO unit, if n is 1 the unit is
on, 0 it is off, e.g. "LIGHT, POWER, 0" would switch the unit
off.
Sending "LIGHT, POWER" without 0 or 1 will report the
R
current status.
The Lumen 200PRO unit must be connected to the plug and
play shutter output of a ProScan III. The Lumen will default to
state 1, Lumen on, if no cable connected.
Version 1.12 June 2017
ProScan® III Manual
Light output
Page 55