Note: Shutters should always be installed with the silvered face facing the
lamp-house.
For more information on installation, please refer to the 'Installation Guides and Help
Sheets' section on the Prior website.
!
NOTE: SHUTTERS MUST ONLY BE RUN AT A MAXIMUM OF 10 HZ

2.34 Fourth Axis or Auxiliary Axis equipment.

Fourth Axis or Auxiliary Axis equipment connects to the 15 way connector, labelled
"AXIS A". Switch off the power using the power switch. Connect and firmly and screw in
the cable both at the controller and the equipment.
The controller will automatically detect Prior accessories and set them up for use. For
non-standard user equipment use the FOURTH commands in the Appendix.

2.35 Connecting the Interactive Control Centre.

Ensure the system is switched off. Plug in the 9 way connector to either the RS232-1
(preferable option) or the RS232-2 connector. Power on the system. The Interactive
Control Centre will detect the ProScan III controller and automatically configure the
internal software. See the Interactive Control Centre manual for further details.

2.36 Installing a Lumen 200Pro

The Lumen 200Pro uses 2 15-way connectors and a round "shutter" connector. The two
15 way connectors should be connected to "FILTER 1" and "FILTER 2"; "AXIS A" can also
be used if necessary. Switch off the power using the power switch. Connect and firmly
screw in the cables for both at the controller and two Lumen 15 way connectors. The
shutter connector is optional and should be connected to the ProScan III "S1, S2 or S3".
Switch off the power using the power switch. Connect and firmly and screw in the cable
both at the controller and Lumen 200Pro connector. If connected the default for the
Lumen200 Pro is OFF. The system must be activated via software – you cannot just switch
on the device itself.
© Prior Scientific Instruments Ltd
Version 1.12 June 2017
Page 12