ProScan® III Manual
Range: Reports the Range of the Piezo.
Advanced TTL control:
This allows the programming of the auto response TTL system. Use this to trigger events
via TTL.
Supported Functions are:
Lumen Attenuator, move to % open.
Filter wheels: move to Next, Previous, Home and Move to position.
Shutter: Open and Close.
XYZ axis: Relative and Absolute Moves.
© Prior Scientific Instruments Ltd
Step Size: Sets the step size for the Step
up and Step down buttons.
Slices: Set the number of slices for the
run slices button.
Delays: Sets the delay between the slices
for the run slices button.
Run Slices: Runs a program stepping
though a number of equal slices between
the position the stage is at to zero. Click
into the area above the green banner
and drag to mouse to make the piezo
stage move. Use the Fine Focus scaling
to increase and decrease movement
speed.
TTL Control Menu:
For basic TTL control and indication use the basic screen,
check the boxes to output high signals and uncheck them to
output low signals. The inputs will indicate if the signal is high
(checked) or low (unchecked).
Version 1.12 June 2017