Prior Scientific ProScan III Manual page 17

Universal microscope automation controller
Hide thumbs Also See for ProScan III:
Table of Contents

Advertisement

Identify where the ancillary box will be controlled. Ensure that the ancillary unit
will be facing in the same direction as the main control box. Remove the
appropriate panel. Slide the case of the ancillary box away from the interior, and
place it next to the main box in the position where it will be installed. You can
ensure that the orientation is correct by noting the position of the slots. If you are
installing it to the LEFT or BOTTOM of the main box, the slots (circled in green)
should be on the side opposite to the main case.
If to the RIGHT or TOP, the slots should be adjacent to the main case. The image
on the next page shows the main and ancillary cases correctly orientated for an
installation on the left. Use the U-clamps provided to fasten the cases together –
note that these may be attached to the ancillary box casing. Then install the panel
you removed from the main casing onto the appropriate open side of the ancillary
casing.
© Prior Scientific Instruments Ltd
Leave the ancillary box as it is.
Slide the case of the main box
away from the interior which will
expose the inner circuit boards.
Version 1.12 June 2017
ProScan® III Manual
Page 15