Warming Up
After turning ON the power supply, allow the product to stand for at least 10 minutes
before use. The circuits are still unstable just after the power supply is turned ON, so
measured values may fluctuate gradually.
A warmup of at least 10 minutes is also required after canceling LD-OFF input if LD-
OFF input is being used.
Sensing Object
The product cannot accurately measure the following types of objects: Transparent
objects, objects with an extremely low reflective sensor ratio, objects smaller than the
beam size, objects with a large curvature, excessively inclined objects, etc.
Mutual Interference
Inserting a calculating unit between amplifier units can prevent mutual interference
between two sensor heads.
Maintenance
• Always turn OFF the power supply before adjusting or connecting/disconnecting
the sensor head.
• Do not use thinner, benzene, acetone or kerosene to clean the sensor head and
amplifier units. If large dust particles adhere to the front filter of the sensor head,
use a blower brush (used to clean camera lenses) to blow them off. Do not blow the
dust away with your mouth. To remove smaller dust particles, use a soft cloth (for
lenses) with a small amount of alcohol. Take care not to wipe them off with exces-
sive force.
Scratches on the filter may cause errors.
ZX2 User's Manual
CONTENTS
INTRODUCTION
PREPARATION
FOR
MEASUREMENT
FLOW OF
OPERATION
BASIC
SETUP
MAIN
APPLICATIONS
& SETTING
METHODS
Height
Steps
and
Warpage
Double
Sheet
Detection
Thickness
Positioning
Eccentricity
and Surface
Deflection
DETAILED
SETTINGS
TROUBLE-
SHOOTING
SPECIFI-
CATIONS
INDEX
SETTING
TRANSITION
CHARTS
13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents