Error message
Cannot start flat frame calibration
because the CCD calibration for High
Resolution mode does not exist.
Please run CCD calibration first.
Option menu - CCD Calibration
Error message
The door was opened during expo-
sure. This image may be inaccurate.
Option menu - CCD Calibration Pro
Error message
Cannot use the flat frame calibration
for the following settings: Light none,
Filter No1, FUJINON VRF43LMD lens
Cannot continue with the automatic
exposure function. The sample may
be out of the photometric range, or
a setting may need to be adjusted.
Help menu
Error message
Failed to open the User Manual.
Please note that a PDF reader (e.g.
Adobe Reader) is needed.
Failed to open End-User Licence
Agreement. Please note that a PDF
reader (e.g. Adobe Reader) is needed.
Failed to open the Getting Started
guide. Please note that a PDF reader
(e.g. Adobe reader) is needed.
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
Meaning
The CCD has not been cali-
brated
Meaning
The door was opened dur-
ing exposure, so the image
may not have been ex-
posed correctly
Meaning
The FlatFrame Calibration
Pro function cannot be
used in chemilumescent
mode
The automatic exposure
function is unable to calcu-
late a correct exposure
time
Meaning
The PDF documents can-
not be opened
6 Troubleshooting
6.6 Error messages and countermeasures
Countermeasure
Perform a flat frame calibration
Countermeasure
Perform the CCD calibration
again
Countermeasure
Use the CCD Calibration option
Change the settings appropri-
ately, or perform an exposure
in Manual mode
Countermeasure
Ensure that the documentation
and PDF reader software is in-
stalled
77