GE ImageQuant LAS 4000 Original Instructions Manual page 70

Hide thumbs Also See for ImageQuant LAS 4000:
Table of Contents

Advertisement

6 Troubleshooting
6.5 Message "ImageQuant LAS 4000 hardware error occurred"
SensKey
SensCode
B001
B002
B004
B008
B010
B020
B040
3
B0 **
except described
above
B101
B102
B104
B108
B110
B120
B140
B1 **
except described
above
A000
4
A100
A200
70
Content
Lifting table error during
startup diagnosis
Filter changer error during
startup diagnosis
Focus error during startup
diagnosis
Iris error during startup diag-
nosis
Epi light source error during
startup diagnosis
UV transluminator error dur-
ing startup diagnosis
White light table error during
startup diagnosis
A combination of errors B004
to B020 above are generated
during startup diagnosis
Lifting table error
Filter changer error
Focus error
Iris error
Transmitted light source error
UV transilluminator error
White light table error
Errors B104 to B120 above
are generated in combina-
tion.
Communication error be-
tween camera head and Im-
ageQuant LAS 4000
Communication retry error
between camera head and
ImageQuant LAS 4000
Countermeasure
Restart the ImageQuant LAS 4000 and Image-
Quant LAS 4000 Control Software
Restart the ImageQuant LAS 4000 and Image-
Quant LAS 4000 Control Software
Getting Started with ImageQuant LAS 4000 28-9607-59 AB