GE ImageQuant LAS 4000 Original Instructions Manual page 73

Hide thumbs Also See for ImageQuant LAS 4000:
Table of Contents

Advertisement

Error message
Components in ImageQuant LAS
4000 have changed.
Please change the exposure method.
Please change settings in
Method/Tray Position window.
Start button - starting exposure
Error message
The interval time must be greater
that 10 seconds when using the In-
crement/Repetition mode.
The maximum exposure time when
using the EtBr method is 10 minutes.
The maximum exposure time when
using the UV light is 10 minutes.
The maximum interval time is 2
hours.
The maximum exposure time is 30
hours.
The maximum interval time when
using the EtBr method is 60 seconds.
The maximum interval time when
using the UV light is 60 seconds.
Please input the file name using the
program settings window.
Please input a valid exposure time.
Cannot start the exposure.
Automatic exposure is only available
when using standard sensivity in the
digitization method.
Automatic exposure cannot be used
at the sensitivity or resolution.
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
Meaning
The default settings have
been changed
Meaning
The time value is inappro-
priate
No file name is given in
Program Settings
The exposure time is not
appropriate
In the Digitize Method, the
Auto function cannot be
used when the sensitivity
is High, Super or Ultra
6 Troubleshooting
6.6 Error messages and countermeasures
Countermeasure
Change the method and/or the
tray position in the
Method/Tray Position dialog
Countermeasure
Enter an appropriate time value
Input a valid filename in the
Program Settings dialog
Choose a suitable exposure
time
Change the sensitivity setting,
or expose an image in Manual
mode