6 Troubleshooting
6.1 Problems with the exposed image
Phenomenon
Unevenness is seen on the im-
age
There is noise on an image ex-
posed using UV trans tray
6.2 Strange sounds or odors are perceived
Phenomenon
The camera head generates a strange
noise
The camera head generates a strange
odor
The camera head emits smoke
ImageQuant LAS 4000 beeps on and
off (four times, at intervals of three
minutes) and the Busy LED blinks
66
Cause
Light source uneven-
ness
A lamp in the UV tran-
silluminator is broken.
The Flat Frame is not
correctly exposed.
Degradation of UV
trans tray
Cause
Camera head failure
Camera head failure
Camera head failure
The door is open
Countermeasure
Check that both left and right Epi light
sources are correctly in place and that
they are of the same type
Turn the power off and contact your GE
Healthcare representative
Expose the Flat Frame again
Purchase a new UV trans tray.
Countermeasure
Turn off the ImageQuant LAS 4000 power
switch immediately and contact your GE
Healthcare representative
Turn off the ImageQuant LAS 4000 power
switch immediately and contact your GE
Healthcare representative
Turn off the ImageQuant LAS 4000 power
switch immediately and contact your GE
Healthcare representative
Close the door. The subsequent processing
should start.