Calibration

Voltage Input Range (50 Volts)

Use the following procedure for calibrating the internal 50 Volt range.
(1)
Switch the instrument on, enter CALIBRATION mode and select INT RANGES from
the CALIBRATION menu as shown on Page 45.
(2)
Select 50V range for calibration:
(3)
Link the voltage input terminals with a short lead and enter the zero point:
(4)
Remove the shorting link from the voltmeter terminals, connect the Vin terminals
of the instrument to a variable voltage source and connect a digital voltmeter
across the source:
49
K415 Issue No. 1