Transmitter Simulator (Tx Sim) Task - GE Druck DPI 610 User Manual

Portable pressure calibrator series
Hide thumbs Also See for Druck DPI 610:
Table of Contents

Advertisement

Taking Measurements
Transmitter Simulation (TX SIM) Task
Provides a current output proportional to the calibrator's measured output pressure
(indicated pressure on indicator only version). Select task TX SIM. Press EXIT to skip set-
up screen if parameters are correct.
TRANSMETTEUR
VALEUR MINIMUM
0.000
4.000
VALEUR MAXIMUM
20.000
20.000
CHANGER
VALEUR
TRANSMETTEUR
VALEUR MINIMUM
0.000
4.000
VALEUR MAXIMUM
20.000
20.000
CHANGER
VALEUR
On completion of TX SIM set-up, the display is configured as follows:
Set-up the output loop as detailed on page 25 and, if necessary, turn on the internal
electrical supply.
To subsequently change any of the TX SIM scaling parameters, press CHANGE VALUE
(F1) to obtain the TX Simulation display.
To change the pressure units, press INPUT and select the required scale units. If the
required scale units are not listed, press SETUP, select SETTINGS and proceed as detailed
on page 38.
K415 Issue No. 1
REGLER VALEUR
kPa
mA
kPa
mA
REGLER VALEUR
kPa
mA
kPa
mA
TASK : TX SIM
SNAPSHOT MODE
PRESSURE INT
CURRENT
CHANGE
VALUE
18
TRANSMETTEUR
VALEUR MINIMUM
0.000
4.000
VALEUR MAXIMUM
20.000
20.000
CHANGER
VALEUR
TRANSMETTEUR
VALEUR MINIMUM
0.000
4.000
VALEUR MAXIMUM
20.000
20.000
CHANGER
VALEUR
24 V OFF
bar
mA
OK
REGLER VALEUR
kPa
mA
kPa
mA
REGLER VALEUR
kPa
mA
kPa
mA