Calibration

Voltage Input Range (5 Volts)

Use the following procedure for calibrating the internal 5 Volt range.
(1)
Switch the instrument on, enter CALIBRATION mode and select INT RANGES from
the CALIBRATION menu as shown on page 45.
(2)
Select 5V range for calibration:
CALIBRATION
PRESSION INT
TENSION
ENTR. COURANT
SORT. COURANT
TEMPERATURE
(3)
Link the voltage input terminals with a short lead and enter the zero point:
(4)
Remove the shorting link from the voltmeter terminals, connect the Vin terminals
of the instrument to a variable voltage source and connect a digital voltmeter
across the source:
SELECTIONNER
SELECT. GAMME
OPTION CAL
47
5 VOLTS
50 VOLTS
SELECTIONNER
OPTION CAL
K415 Issue No. 1