Advanced Task

Flow Function

To apply the flow function to a selected input, enable FLOW from the process
menu and press ENTER. The square root symbol is displayed beside the input
value to indicate that the FLOW function is active:
FCT. : CONFIG.
MODE ENREG. RAPIDE

%Span

To convert a selected input display from a numerical value to a percentage of
full-scale reading, enable SPAN from the process menu and press F1 (SETTINGS)
to provide access to the span definition parameters. Two span definitions are
required, Zero and Full Scale.
To leave span at current setting, press EXIT.
To define zero and full-scale settings, proceed as follows:
DEFINE ECH.
VALEUR AFFICHEE
0% =
100% =
50.000
ENTER UNE VALEUR OU F2
POUR SELECTIONNERVALEUR
ACTUELLE OU EXIT POUR
QUITTER SANS MODIFICATIONS
CHANGER
VALEUR
To cancel %SPAN, press INPUT and turn function OFF at the process menu.
K415 Issue No. 1
24V OFF
V
bar
SELECTIONNER
VALEUR
V
_
V
LECTURE
ACTUELLE
To cancel FLOW, press INPUT and turn function
OFF at the process menu.
DEFINE ECH.
VALEUR AFFICHEE
0% =
0.000
_
100% =
ENTER UNE VALEUR OU F2
POUR SELECTIONNERVALEUR
ACTUELLE OU EXIT POUR
QUITTER SANS MODIFICATIONS
CHANGER
VALEUR
24
Process Functions
SELECTIONNER
VALEUR
V
V
LECTURE
ACTUELLE