Calibration

Voltage Input Range (5 Volts)

Use the following procedure for calibrating the internal 5 Volt range.
(1)
Switch the instrument on, enter CALIBRATION mode and select INT RANGES from
the CALIBRATION menu shown on page 43.
(2)
Select 5V range for calibration:
(3)
Link the voltage input terminals with a short lead and enter the zero point:
(4)
Remove the shorting link from the voltmeter terminals, connect the Vin terminals
of the instrument to a variable voltage source and connect a digital voltmeter
across the source:
SELECT CAL OPTION
CALIBRATION
PRESSURE INT
VOLTAGE IN
CURRENT IN
CURRENT OUT
TEMPERATURE
+
Cal V
-
DVM
SELECT RANGE
5 VOLTS
30 VOLTS
max 30V
Vin
55mA
max
mA in
mA Sink
45
SELECT CAL OPTION
K0430 Issue No. 2