Ma Step - GE Druck DPI 610 User Manual

Portable pressure calibrator series
Hide thumbs Also See for Druck DPI 610:
Table of Contents

Advertisement

Advanced Task

mA Step

To select one of the electrical output programs, press the OUTPUT key and proceed as
follows:
SELECT SORTIE
PRESSION INT
PRESSION EXT
OUTPUT
PAS mA
RAMPE mA
VALEUR mA
On selection of (e.g.) Linear, the output display window changes to show the selected
program of output currents:
For current sourcing applications, turn on the 24V supply as shown on page 28. For
current sinking applications, connect an external supply as shown on page 25 and
leave the 24V supply switched OFF.
Press RUN (F1) to run program. A flashing status display CHECK LOOP indicates a
fault in the external loop i.e. supply fault or open circuit.
Note:
The dwell time at each step is approximately 10 seconds.
Press STOP (F1) when running to stop at any point. Press RUN (F1) to resume.
K415 Issue No. 1
SELECTIONNER
SORTIE
INT
UNITES
24V OFF
FCT. : CONFIG.
MODE ENREG. RAPIDE
bar
LINEAIRE 4-20 mA PAS
4
8
12
16
20
mA
ARRET
DEMARRER
26
Select Output
PAS mA
LINEAIRE
4-20
mA
LINEAIRE
0-20
mA
RAC CARR
4-20
mA
RAC CARR
0-20
mA
VANNES
4-20
mA
DCS
4-20
mA
FCT. : CONFIG.
24V OFF
MODE ENREG. RAPIDE
bar
LINEAIRE 4-20 mA PAS
4
8
12
16
20
mA
EN COURS
ARRET
SELECTIONNER
PROGRAMME