Relief Valve Test (Rel Valve) Task - GE Druck DPI 610 User Manual

Portable pressure calibrator series
Hide thumbs Also See for Druck DPI 610:
Table of Contents

Advertisement

Taking Measurements

Relief Valve Test (REL VALVE) Task

To carry out a relief valve test, press TASK and select REL VALVE. Connect the output
pressure port of the instrument to an external system as shown below:
A
B
A - External pressure source (indicator instruments only)
C - Relief valve under test
To change the pressure units, if required, press INPUT and select the required
units using the cursor keys.
If necessary, turn on the 24 Volt internal supply by pressing OUTPUT, select 24
VOLT and switch ON with the right cursor button and press ENTER.
Close the vent valve and, using the hand-pump or external pressure supply, apply
pressure to the relief valve under test.
When the relief valve operates, the maximum recorded pressure indicates the
operating point of the valve.
Note: The STORE key can be used for this purpose. Use right cursor key initially,
followed by up/down keys to enter Snapshot text).
Record the test results.
Open vent valve to release test pressure.
Note: If using external pressure supply, isolate supply before opening the vent
valve.
DPI 615
FCT. : SOUPAPE
MODE ENREG. RAPIDE
MAXIMUM
MINIMUM
RESET
MAX/MIN
F1
F2
B - Pressure regulator
19
C
24V OFF
bar
bar
K415 Issue No. 1