Memory Operations
Running Procedure Files (DPI 615 instruments only)
To run a procedure, make sure the instrument is set to the Store mode, Data Log (see
page 37), and proceed as follows:
MEMORY USED
FEEDPRESS11
G3459
RECALL
TEST1
TEST2
TEST5
FILE 1 of 9
DATE: 04/11/99 TIME 10:18
After selecting F1, proceed by entering the User ID and Serial Number and then select F1
(Continue) and follow the on-screen procedural instructions:
..........
FEEDPRESS11
GO TO
0.000 bar
0.04
%span PASS
LOG
NOW
SETUP & LOG POINT 1
When the test procedure for a given UUT has been completed, the result of running the
first test is stored as an AS FOUND file. This file cannot be overwritten. Any subsequent
tests on the device are stored as an AS LEFT file that is overwritten each time the
procedure on this device is run.
When recalling the results of a procedure, the choice of AS FOUND or AS LEFT is provided
(refer to page 36).
ERASE
24V OFF
FEEDPRESS11
GO TO
V
DATA POINT LOGGED
bar
POINT 1 STORED
SELECT FILE
FEEDPRESS11
DEVICE UNDER TEST
I/P LOW:
I/P HIGH
O/P LOW:
O/P HIGH
MAX PRESS:
POINTS:
MAX. ERROR:
PROCEDURE
..........
24V OFF
0.000 bar
V
%span PASS
bar
35
Data Log
0 BAR
20 BAR
0 VOLTS
10 VOLTS
22
5 UP/DOWN
0.1% SPAN
RUN
REVIEW
LOGGED FILE
..........
FEEDPRESS11
24V OFF
GO TO
5.000 bar
0.04
%span PASS
bar
LOG
NOW
SETUP & LOG POINT 2 etc.
K415 Issue No. 1