Calibration
(4)
Measure the 10% full-scale output current and enter the value measured on the
external milliammeter:
(5)
Measure the 90% full-scale output current and enter the value measured on the
external milliammeter
YES
(6)
Press the ENTER key to accept the calibration. Press the EXIT key four times to
quit the CALIBRATION and SETUP modes.
(7)
Verify the instrument's output current calibration by setting up the loop test
currents as shown in Table 4 and checking the set values against the
milliammeter standard.
(8)
Press the OUTPUT key, select 24 VOLT and turn the 24V supply to ON. Proceed as
follows:
K415 Issue No. 1
Store Cal point?
NO
Measure 90% FS - output
Press ENTER key when
applied value is stable
54
Enter Measured Current:
:
RUN