Emerson Rosemount 8800D Series Reference Manual page 181

Vortex flowmeter
Hide thumbs Also See for Rosemount 8800D Series:
Table of Contents

Advertisement

Electronics Verification
May 2016
Example 2 (SI units)
Fluid
Line size
Line pressure
Operating
Temperature
Viscosity
Input
Frequency
K-factor
(compensated)
M
Therefore, an input frequency of 650 Hz represents a flow rate of 3446.4 kg/hr in this
application.
For a given input frequency, you may also determine the electrical current output. Use the
prior example with an input frequency of 275 Hz:
URV= 3600 kg/hr
I = 10.48 mA
Therefore, an input frequency of 275 Hz represents an output electrical current of 10.48
mA.
Electronics Verification
= Saturated Steam
= 80 mm
= 700 kPas
= 170 °F
= 0.015 cp
= 650 Hz
= 10.715 pulses/gallon (via Field Communicator or AMS Device Manager)
= F(Hz) /(K x C
16
= 650/{10.715 x (C
= 650/{10.715 x (0.0733811/4.169)}
= 650/(10.715 x 0.017602)
= 3446.4 kg/hr
(
×
) LRV
F
K
C
16
I
=
---------------------------------------------------- -
URV LRV
(
275
10.715 0.017602
I
=
-----------------------------------------------------------------------------
3600 0
URV
= 3600 kg/hr
LRV
= 0 kg/hr
C
= C
16
11
ρ
Density(
)
= 4.169 kg/cu-mtr (operating)
)
ρ
/
)}
11
LRV= 0 kg/hr
×
(
)
16
+
4
×
) 0
×
(
Reference Manual
00809-0100-4004, Rev DC
ρ
/
(from
Table C-1 on
page168)
F
= 275 Hz
in
)
16
4
+
173