Thermo Scientific AI 1310 User Manual page 96

Table of Contents

Advertisement

4
AI 1310/AS 1310 Control
Guidelines for Programming with Different Injectors
Table 11. Summary of Injection Parameters and Conditions
Mode
Technique
Split
Hot Needle
Split
Cold Needle
Splitless
Hot Needle
Splitless
Cold Needle
1
SSL Nano-Volumes
Cold Needle
1
Only with 0.5μL syringe.
72
AI 1310/AS 1310 Sampling System - User Guide
Column
Liner Type
Insertion
40 mm
Tubular Liner
(3 or 5 mm ID)
40 mm
Tubular Liner
(3 or 5 mm ID)
64 mm
Tapered Liner
(3 or 5 mm ID)
50 mm
Tapered Liner
(3 or 5 mm ID)
40 mm
Double Tapered Liner No
Packing of
AI 1310/AS 1310 Injection Parameters
Liner
Not required
Injection depth = Standard
Pre Inject Delay Time = 3-5 s
Post Inject Delay Time = 0 s
(Chromeleon CDS)
Pre Injection Dwell Time = 3-5 s
Post Injection Dwell Time = 0
(Xcalibur, ChromCard, ChomQuest CDS)
Packing material
Injection depth = Minimum
at half height
Pre and Post Inject Delay Time = 0 s
(Chromeleon CDS)
Pre and Post Injection Dwell Time are
automatically inhibited
(Xcalibur, ChromCard, ChomQuest CDS)
Not Required
Injection depth = Standard
Pre Inject Delay Time = 3-5 s
Post Inject Delay Time = 0 s
(Chromeleon CDS)
Pre Injection Dwell Time = 3-5 s
Post Injection Dweel Time = 0
(Xcalibur, ChromCard, ChomQuest CDS)
Packing material
Injection depth = Minimum
at bottom
Pre and Post Inject Delay Time = 0 s
(Chromeleon CDS)
Pre and Post Injection Dwell Time are
automatically inhibited
(Xcalibur, ChromCard, ChomQuest CDS)
Injection depth = Minimum
Pre and Post Inject Delay Time = 0 s
(Chromeleon CDS)
Pre and Post Injection Dwell Time are
automatically inhibited
(Xcalibur, ChromCard, ChomQuest CDS)
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

As 1310

Table of Contents