Figure 33. Assembling on the Back Injector Module (2)
4. Place the L-shape bottom plate on the GC top and guide the three fixing screws into the
corresponding fixing holes. See A', B', and C'. Tighten the three fixing screws.
5. If not already done, mount and fix the guide pivot on the top plate by using the proper
fixing screw.
6. Place the top plate over the L-shape bottom plate.
7. Guide the retainer and its fixing screw into the sliding slot present on the top plate.
8. Align the retainer fixing screw into the corresponding fixing hole located on the L-shape
bottom plate.
9. Tighten the retaining fixing screw.
10. Clamp the two plates guiding the second retainer and its fixing screw into the hole
provided on the L-shape bottom plate.
11. If you must install an AS 1310, screw the support pin into the dedicated hole of the top
plate.
Note The support pin is NOT required if you must install an AI 1310.
Installing the Sampling System Support on the TRACE 1300/1310 GC
Guide Pivot
Retainer
Retainer
A'
C'
AI 1310/AS 1310 Sampling System - User Guide
Support Pin
L-shape Bottom Plate
B'
3
Installation
37