Table 3. Hardware Setup Comparison Table
Chromatography Data System
(CDS)
Chromeleon 7.2
Xcalibur
Fundation 3.1; GC driver 3.X
TraceFinder 3.2
Chrom-Card 2.12
ChromQuest 5; SP2015 R2
EzChrom 3.22; SP2015 R2
1
The unplugging of the power cord from the unused AI 1310/AS 1310 autosampler could be required
2
Not Applicable
To perform the HWSetup 1 of two AI/AS 1310 in Gemini Configuration
1. Using the cable provided (PN 23043672), connect the 4-pin male connector marked
TWIN SYNC, on the back of the sampling unit of the sampler "a", to the 4-pin male
connector marked TWIN SYNC on the back of the sampling unit of the sampler "b".
2. Using the Gemini Y shape cable provided (PN 23043623), connect:
a. the 6-pin main connector labelled GC to the connector marked AUTOSAMPLER
HANDSHAKE on the back of the GC.
b. the first branch of the cable, labelled AI/AS 1, to the 6-pin female connector marked
GC on the back of the sampling unit of the sampler "a".
c. the second branch of the cable, labelled AI/AS 2, to the 6-pin female connector
marked GC on the back of the sampling unit of the sampler "b".
3. Using the RS232 cable (PN 23043453) provided in the standard outfit of each AI/AS
1310, connect:
a. the 9-pin connector marked RS232 of the first sampler to a PC serial port (e.g. COM
1), or to an USB port, interposing an USB-RS232 adapter not provided.
b. the 9-pin connector marked RS232 of the second sampler to a second PC serial port
(e.g. COM 2), or to an second USB port, interposing a second USB-RS232 adapter
not provided.
4. Connect the power cable to the AC Input connector (Mains socket) on the autosampler
and to the wall outlet.
Installation of the AI 1310/AS 1310 on the GC
High Throughput
Single
Mode
Mode
HWSetup 1
HWSetup 1
See
Figure 49
See
Figure 49
2
NA
HWSetup 2
See
Figure 50
2
NA
HWSetup 2
See
Figure 50
Alternative
HWSetup 2
See
Figure 50
See
Figure 50
HWSetup 1
HWSetup 1
See
Figure 49
See
Figure 49
HWSetup 1
HWSetup 1
See
Figure 49
See
Figure 49
AI 1310/AS 1310 Sampling System - User Guide
3
Installation
Confirmation
1
Mode
HWSetup 2
See
Figure 50
1
HWSetup 2
See
Figure 50
1
HWSetup 2
See
Figure 50
1
HWSetup 2
See
Figure 50
HWSetup 1
See
Figure 49
HWSetup 1
See
Figure 49