Thermo Scientific AI 1310 User Manual page 65

Table of Contents

Advertisement

Thermo Scientific
Figure 37. Assembling (2)
Top Plate Back not for
155-position Sample Tray
Guide Pivot
Top Plate Front
Support Pin
3. Guide the five fixing screws into the corresponding fixing holes located on the GC top.
See A', B', C', D', and E'. Tighten the five fixing screws.
4. If not already done, mount and fix the guide pivot on each top plate by using the proper
fixing screw.
5. Place the two top plates over the double bottom plate. Guide each retainer and its fixing
screw into the sliding slot present on each top plate.
6. Align each retainer fixing screw into the corresponding fixing hole located on the double
bottom plate.
7. Tighten each retainer fixing screw to block the corresponding upper plate to the double
bottom plate.
8. Clamp the corresponding plates guiding the second retainer and its fixing screw into the
hole provided on each L-shape bottom plate.
9. Screw the support pin into the dedicated hole of each top plate.
Note The support pin is NOT required if you must install an AI 1310.
Installing the Sampling System Support on the TRACE 1300/1310 GC
Guide Pivot
Retainer
Retainer
A'
D'
AI 1310/AS 1310 Sampling System - User Guide
Support Pin
Retainer
C'
E'
Double Bottom Plate
B'
3
Installation
41

Hide quick links: