Setting the Combination Trigger

IM 701450-51E
24.
Press the Combination soft key. The Combination menu appears.
25.
Press CAN Only, CAN on Pattern, or CAN -> Pattern soft key.
Select CAN Only to activate a trigger only on the trigger conditions of the CAN Bus
signal. Select CAN on Pattern to activate a trigger when the trigger conditions of the
CAN Bus signal are met while the trigger conditions of CH2 to CH8 (CH2 to CH4 on the
DL7440) are met. Select CAN -> Pattern to activate a trigger when the trigger
conditions of CH2 to CH8 (CH2 to CH4 on the DL7440) are met after the trigger
conditions of the CAN Bus signal are met.
When CAN Only Is Selected
26.
Press the Set Pattern soft key. The Set Pattern dialog box displays the
assignment conditions of the CAN Bus signal of CH1.
27.
Press ESC. The Set Pattern dialog box closes.
Proceed to step 32.