1.7

Error Messages

Code
Message
37
Aborted the analysis.
38
Data not detected.
Execute again after changing the settings or
reacquiring the waveform.
730
Pattern is not specified.
739
Analyzed data does not exist.
Execute the analysis.
870
Cannot be specified.
Invalid byte or bit.
IM 701450-51E
A message may appear during operation. This section describes the meanings of the
messages and their corrective actions. This section lists only the error messages related
2
to the I
C Bus signal analysis function. There are other error messages related to the
DL7440/DL7480 and communications. These messages are described in the DL7440/
DL7480 User's Manual (IM701450-01E) and the DL7440/DL7480 Communication
Interface User's Manual (IM701450-17E) .
You can set the messages to be displayed in English or Japanese. For the procedure of
setting the message language, see section 15.1, "Changing the Message Language and
Turning ON/OFF the Click Sound" in the DL7440/DL7480 User's Manual (IM701450-
01E) .
If the corrective action requires servicing, contact your nearest YOKOGAWA dealer for
repairs.
Action
Enable Address Pattern or Data Pattern in the
Search Setup dialog box.
Execute the analysis.
Increase the number of valid bytes.
1
Page
1-27
1-4, 1-10,
1-25, 1-30
1-29
1-27, 1-31
1-6, 1-11
1-37

Advertisement