2
Chapter 1
I
C Bus Signal Analysis Function
1.1

Overview of the I

Function
2
About the I
C Bus Signal Analysis Function
Trigger Function <See page 1-4 for the operating procedure>
Address & Data Trigger Example
• When Activating a Trigger on the Start Condition or Address Pattern/Data Pattern
Start condition
SDA
SCL
Non-ACK Trigger Example
• When Activating a Trigger When the Acknowledge Bit Is Not Present (When the SDA Signal is high)
Start condition
SDA
SCL
Analysis Function <See page 1-25 for the operating procedure>
Search Function <See page 1-28 for the operating procedure>
IM 701450-51E
2
C Bus Signal Analysis
2
I
C Bus is an abbreviation for Integrated Circuit Bus. It is a bidirectional bus for
connecting ICs. By using this function, you will be able to analyze data while displaying
2
the I
C-bus signal waveform. The I
following three main functions.
Various trigger conditions for the I
condition is detected, when an Acknowledge bit is not detected, when the specified
_
address (7 bits + R/W
) pattern is met, or when the data pattern is met or not met (Non-
Ack trigger or address & data trigger).
Triggers can also be activated by combining the I
signal (CH2) and the CH3 to CH8 signal (CH3 and CH4 on the DL7440) (combination
trigger).
Acknowledge bit
Address + R/W bit
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
Acknowledge bit
Address + R/W bit
Acknowledge present (L)
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
2
Analyzes the I
C Bus signal that was acquired using the trigger function and lists the
analysis data (hexadecimal notation) at the byte level and the status of the Acknowledge
bit. When you select any analysis data on this analysis result list, the I
waveform corresponding to the analysis data is automatically displayed. You can view
the details of the analysis result such as the position of the start condition/stop condition
or the time from the analysis reference point on the detailed analysis list. You can save
the detailed analysis list on a storage medium in ASCII format.
In addition, you can analyze the data by switching between two sets of I
(CH1: SCL and CH2: SDA or CH3: SCL and CH4: SDA).
2
After analyzing the I
C Bus signal that was acquired using the trigger function, this
function searches in the forward or reverse direction for data that matches a specified
address pattern, data pattern, or Acknowledge bit condition, and displays the matched
data expanded on the ZOOM display. You can specify the address pattern or data
pattern using binary or hexadecimal values. You can also search indefinite data.
2
C Bus Signal Analysis Function consists of the
2
C Bus signal can be set up such as when a start
2
C Bus signal (SCL signal (CH1)/SDA
Data byte
...
...
Trigger activated here
Data byte
Acknowledge not
present (H)
...
...
Trigger activated here
Stop condition
Stop condition
2
C Bus signal
2
C Bus signals
1-1