PROGRAMMING THE EXPOSURE VALUES

10.1 Programming the default exposure and density values

Programming the density values
User's manual
NOTE Make sure, that no exposure region is selected, i.e. no
NOTE The exposure parameters - time and kV - are program-
0 . 1 6
s
SELECT
c . 1 6
s
8 mA
6 3
SELECT
kV
d 1
s
SELECT
c - 4
s
PROGRAMMING THE EXPOSURE VALUES
The default exposure values can be programmed for both
the adult and child mode. When the indicator light is on,
the unit is in the child mode. The current exposure values
are shown on the time and kV displays.
preprogrammed setting indicator light is on.
med corresponding to the normal mode density value 0.
The time value will be automatically changed according
to the selected density value in both the normal and digi-
tal imaging mode when you exit the programming mode.
Press and hold down the select key until you have heard
two signal tones (about 4 seconds) to enter the program
mode. The time display will start to flash, the default ex-
posure values will appear on the displays and the ready
light will start to flash.
The exposure time value is changed with the parameter
adjustment keys. The exposure times are shown in section
12 "EXPOSURE VALUE TABLES" on page23.
Press the select key briefly, the kV display will start to
flash and the exposure time value is stored in the memory.
The kV value can now be changed with the parameter ad-
justment keys.
When the select key is pressed a second time briefly, the
current density value starts to flash on the time display and
the kV value is stored in the memory.
Prostyle Intra X-ray Unit 19