GP Monitor
3.4.12 Factory Defaults
If you need to reset the GP monitor to its factory default settings, you
can do so by pressing and holding the
the
button 3 times.
Parameter
Emissivity
Sensing Head/Device
type
Display Resolution
Setpoints (SP1, SP2)
Peak hold
Valley hold
Averaging
Ambient Temperature
Compensation (t-amb)
Offset
Table 10: Factory-set Default Values
3.5 Quick Reference
On the following pages are quick reference charts for setting up the
GP Monitor with sensing head and input devices and for adjusting
the monitor's normal and advanced functions.
34
Value
Setting Range
0.95
0.1 - 1.09
H001
H001 - H007
0
0 - 4
off
According to Sensing
Head/Device type
0-998 s (999 s = ∞)
0
0-998 s (999 s = ∞)
0
0 – 60 s
0
0
-18 to 1200°C
(0 to 2200°F)
0
-50 to 50°C
(-99 to 99°F)
Rev. F 10/2011
button and then pressing
Remarks
For 0 - 5 V, 4 - 20 mA and
thermocouple inputs,
emissivity is not functional
see section 3.4.4, page
23
∞: reset by hardware
∞: reset by hardware