Sokkia SRX1 Operator's Manual page 159

Series srx surveying instruments
Table of Contents

Advertisement

PROCEDURE Check
1. Find an area of flat ground where two points
100m apart can be selected.
Set up the Instrument at point A and the reflective
prism at point B. Establish a point C half way
between points A and B.
2. Precisely measure the horizontal distance
between point A and point B 10 times and
calculate the average value.
3. Place the SRX at point C directly between points
A and B and set up the reflective prism at point A.
4. Precisely measure the horizontal distances CA
and CB 10 times each and calculate the average
value for each distance.
5. Calculate the additive distance constant K as
follows.
K = AB - (CA+CB)
6. Repeat steps 1 to 5 two or three times.
If the additive distance constant K is within ±3mm
even once, adjustment is unnecessary.
If it always exceeds this range, have your Sokkia
service representative perform an adjustment.
23. CHECKS AND ADJUSTMENTS
151

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Srx2Srx3Srx5

Table of Contents