• By inputting the temperature and air pressure values, the atmospheric correction value is
calculated and set into the memory. Calculate the atmospheric correction factor as shown in the
following formula.
ppm = 282.59 -
1 + 0.003661 x air temperature (°C)
"30.2 Atmospheric Correction for High Precision Distance Measurement"
•If the weather correction is not required, set the ppm value to 0.
PROCEDURE Recording and editing target information
The [LIST] softkey is be displayed when either "Reflector" or "Prism const." is selected in the EDM tab
of <EDM configurations>.
1. Press [LIST] to display a list of all recorded
targets.
• [ADD]: Displays <Reflector list>. Select the
desired target from this list and press [OK] to
add to the list in <Reflector setting>. Up to a
maximum of 6 targets can be recorded.
• [DEL]: Deletes the selected target.
0.2942 x air pressure (hPa)