23.4
Collimation
With this option you can measure collimation error in your instrument so that the SRX can correct
subsequent single face observations. To measure the error, make angular observations using
both faces.
PROCEDURE
1. Select "Inst. cons." in <Configuration>.
2. Select "Collimation".
3. Sight the reference point in Face 1 and press
[OK]. Telescope rotates and vertical circle is
indexed.
•
Do not look through the telescope eyepiece
while the motor drive is in operation. An eye
could be struck by the telescope and cause
injury.
4. Sight the reference point in Face 2 and press
[OK].
23. CHECKS AND ADJUSTMENTS
143