A`. Measurement Procedure: Ic-Vceo_R Application Test without series collector
resistor.
- Starting from pre-defined test setup of My favorite Setup
This test approach starts measurements by using a pre-defined test setup saved in
My favorite Setup instead of starting from a scratch by using an Application Test
Library.
Setup and execute the test by following the next steps. The same instruction steps
are illustrated by the numbers on the "1`. Starting from Application Test Library"
side of figure 3-13.
Step 1. Click the Preset group of My Favorite Setup.
Step 2. Select Example_AT preset group.
Step 3. Select Ic-Vceo_R0 (Click the Ic-Vceo_R0).
Step 4. Click "Recall" button.
Then go to step 5 of "Starting from Application Test Library" above, and continue by
following the test step numbers.
60