side as shown in figure 2-8 (reside in main section).
For disconnecting the interlock cable, hold the metal part and then pull the
connector by turning it.
Step number 3:
Using a 16493L GNDU Cable, connect the GNDU on the B1505A to the GNDU1
Input on the N1259A.
Step number 4:
Using two 16494A Triax Cables, connect the Force and Sense connectors on the
lower B1510A HPSMU (SMU1) to the respective connectors on the HPSMU1
input of the N1259A.
Note:
HPSMU uses a pair of two triax cables. HCSMU cables looks similar as
HPSMU cables, but HCSMU cables are a pair of a coax cable and a triax cable.
It is a good practice for distinguishing these two cables.
Step number 5:
Using two 16494A Triax Cables, connect the upper B1510A HPSMU (SMU2) to
the HPSMU2 on the N1259A.
Step number 6:
Using a N1300A CMU Cable, connect the B1520A CMU to the respective connec-
tors of MF CMU Input (Hcur, Hpot, Lcur, Lpot) on the N1259A.
Note: Leave the green cable with a round terminal as it is.
Step number 7:
Using a 16493S HCSMU Cable, connect the Force and Sense connectors on the
B1512A HCSMU to the respective connectors of HCSMU1 Input on the N1259A.
Note:
HCSMU cables are a pair of a coax cable and a triax cable. In case of
HPSMU, it uses a pair of two triax cables. It is a good practice for distin-
guishing these two cables.
Step number 8:
Using a 16493T HV Triax Cable, connect the Force connector on the B1513A
HVSMU to the HVSMU1 input of the N1259A
Tips:
Make sure to screw in the connector firmly by hand to the end point where
you feel some resistance.
Step number 9:
Connect power cable to the fixture.