A2-2. Cable Connection between the B1505A and the N1259A Test Fixture
Before starting the measurements, connect the cables between the B1505A and the
N1259A as shown in figure A2-3.
The breakdown of each step with cable figures and the connector locations are
shown in figure A2-4.
These connections are used for all the measurement examples, and there is no need
for changing this configuration.
Connections for 1x HCSMU Configuration (20 A) with Opt 300 Module selector
Back side of the B1505A
B1513A
HVSMU
Triax
B1512A
HCSMU
B1520A
MFCMU
B1510A
HPSMU
B1510A
HPSMU
GNDU
Figure A2-3. Total connection setup for the measurement with the Module Selector.
Step number1:
Step number 2:
108
8
Coax
7
6
5
4
3
2
1
Using a 16493G Digital I/O Cable, connect the Digital I/O connector on the
B1505A to the Digital I/O connector on the N1259A test fixture.
Using a 16493J Interlock Cable, connect the Interlock on the B1505A and the
Interlock on the N1259A
Tips:
For connecting the interlock cable, hold the black plastic part and then turn
the connector by pressing toward the interlock connector in the instrument
Back side of the N1259A
Leave it open
Coax
Triax
9