IMPORTANT
Before starting a measurement, check the settings for the module selector. If the
setting is different from the figure A2-5 setup, please change the setup as in the
figure.
You can view and change the configuration of the module selector as follows.
1.Click the configuration button
2.Select the Module selector tab.
3. Press "Close "
With this setup, the N1259A Status LED displays the current connection status.
Figure A2-5. Module Selector setup for the measurement with the Module Selector.
a. Check "Enable Module Selector"
b. Check "Auto Detection"
c. Uncheck "Enable Series Resistor"
Tips:
There is no indicator on the N1259A test fixture for showing this status.
Therefore this checkbox should better be set to off, every time just after the
use of this function, for preventing any unexpected test result in later use of
the B1505A.
d. Check the SMU modules
e. After modification, press "Apply" button.
a
c
b
on the right side of the screen.
d
e