Section 2
Operation

Check Fuse

Change Configuration

2-10
Orbital Shaker
Check Fuse alerts the operator that there
is no communication with the motor
drive circuit board (historically the
primary drive motor fuse has blown). In the alarm state, the audible alarm
is muted by pressing the Silence button, but rings back in about 30
minutes, if the alarm condition is still present. However, the message and
indicator lights persist until the fault is corrected. When the unit is turned
on after fuse replacement, all alarm indicators are automatically cleared.
Warning Fuse replacement must be performed by qualified service
personnel. See Service section.
To access the system Configuration menu, press the down arrow, the up
arrow and the Silence button in that sequence.
This screen will appear on the
display.
Pressing the down arrow continues
with system configuration.
Pressing the up arrow returns to the Operating Screen.
During the following configuration procedures, menu options are given to
either modify a setting as it appears in sequence, or scroll past to the next
item. If no selection is made by pressing a button or arrow, the display will
revert to the Operating Screen in about 15 seconds. The complete
configuration menu is shown in the chart at the end of this section.
Note In these procedures, values and settings for time, temperature, speeds,
alarms, and so forth are shown on the display screens. These numbers are
for example only and may not be the values encountered when
programming your unit.