Section 1
Installation

Installing the Platform

1-4
Orbital Shaker
Note If your platform is already installed, skip this section and continue to
next section.
Caution Remove the shipping bracket and install the shaker platform
before plugging in or attempting to operate the unit.
After removing the orbital mechanism shipping bracket and installing the
platform, remove this protective decal from the control panel to begin
shaker operation.
IMPORTANT USER INFORMATION
WARNING: DO NOT START
SHAKER W/O FIRST REMOVING
THE PLATFORM HOLD DOWN
BRACKET AND INSTALLING
THE PLATFORM.
Figure 1-6. Warning Decal
To protect the shaker's orbital mechanism during shipment, a sheet metal
shipping bracket (Figure 1-7) is installed and must be removed before the
unit can be operated. Using a 7/16" and 9/16" hex wrench, remove the
three 1/4" and the single 3/8" screws. Retain this hardware for future
shipping.
This shaker accommodates either a 5/16" nominal heavy-duty, 29.5" wide
x 18" front-to-back platform.
All shaker platforms are attached to their orbital mechanisms with six 1/4-
20 hex socket flathead screws Grade 8. These screws are hardened and
should not be exchanged with any other screw type. The 5/32" hex socket
wrench, included with the shaker, must
be used when attaching the platform.
Refer to Figure 1-8.
Figure 1-8. T-Handle Wrench
and Hex Socket Head Screw
Caution Do not attempt to use a Phillips
s
head screwdriver.
s
Figure 1-7. Shipping Bracket