Chamber Drain; Condensate Drain; Foot Pedal - Thermo Scientific MAXQ SHKE435HP Operating And Maintenance Manual

High performance console incubated orbital shaker
Table of Contents

Advertisement

Section 1
Installation
Figure 1-4. Chamber Drain Hose Location

Chamber Drain

A drain is provided in the bottom of the chamber for convenience when
cleaning or removing spills (Figure 1-4). A clear vinyl hose and plastic
valve is connected to the drain and accessed by removing the front grille
assembly and the lower front panel. The grille is removed by gently pulling
it off. It is held in place by six push-in type retainers.
To remove the cabinet panel located in back of the grille, remove the six
Phillips screws; three on the top and three on the bottom. It may also be
necessary to loosen the two left side Phillips screws which hold the foot
pedal assembly to the shaker frame.

Condensate Drain

A 1/4" stainless steel condensate drain
is located on the back of the shaker to
remove any water that may collect in
the air ductwork. Refer to Figure 1-5.
Figure 1-5. Drain Location

Foot Pedal

The lid of the console shaker is counterbalanced for ease of opening and
closing. A foot pedal on the lower right of the front of the cabinet is also
provided for operator convenience and ready access to the chamber.